PermeH2O
Water Vapor Permeability Tester
1 OR 2 CELLS
Permeability tester with integrated controls for temperature and relative humidity.
- Wide measuring range
- Long-life detectors
- Upgradeable with:
- O2 sensor
- CO2 sensor
- Fully software driven
- Barometric compensation
- Graphic software interface
- Automatic humidity adjustment
- No sample cutting required
- Package analysis accessory for bottles, etc.
NORMS: PermeH2O complies with ASTM F1249, TAPPI T557, JIS K-1729, ISO 15106-2 Norms.
PermeH2O
Water Vapor Permeability Tester 1 or 2 CELLS Permeability tester with integrated controls for temperature and relative humidity. Wide measuring range Long-life detectors Upgradeable with: O2 sensor CO2 sensor Fully software driven Barometric compensation Graphic software interface Automatic humidity adjustment No sample cutting required Package analysis accessory for bottles, etc.
NORMS: PermeH2O complies with ASTM F1249, TAPPI T557, JIS K-1729, ISO 15106-2 norms.
PermeH2O is at the forefront of measuring water vapor transmission rate (WVTR) on a wide range of materials, including mono- and multilayer plastic films, metallized films with special surface coatings, laminates, and co-extrusions. These materials are extensively used in the food, beverage, pharmaceutical, and electronics industries.
The dynamic-differential measurement technique employed by PermeH2O allows for the discrimination of lateral losses due to potential surface imperfections of the sample, a capability not achievable with other analysis techniques.
The advanced PermeH2O ExtraSolution® software empowers operators with autonomy after minimal training. Key parameters such as temperature, relative humidity, and sample conditioning times are pre-set upon software initiation, minimizing human intervention and ensuring result accuracy.
Throughout the measurement process, the software maintains constant monitoring and precise recording of conditions, presenting data both textually and graphically over time. Measurement conditions are saved in PDF and text formats for maximum compatibility with other systems.
Electronic control of gas flows renders PermeH2O insensitive to line pressure variations and ambient conditions. The single-point cell sealing system eliminates the need for sample cutting, streamlining measurement procedures and ensuring reliable, reproducible results.
Test range H2O | 0.002-100 g·m-2·24h-1bar-1 (unmasked) 0.04-2000 g·m-2·24h-1bar-1 (masked) |
Test Sample Size | 50 cm2 – About 2,5 mm thickness max |
Test temperature range | 10-50° C |
Relative humidity H2O (wet side) | 5-95% |
Carrier flow (N2) | 10-36 ml/min automatically controlled |
Carrier gas | N2 5.0 |
N2 pressure | 2.0 bar |
Gas connections | 1 Standard Ham-Let 1/8″ |
Software | LabView based with USB interface |
PC with preinstalled software | WindowsTM OS and LCD monitor |
Power supply | 100-220 AC, 50-60 Hz |
Apparatus size (cm) | 41 W / 61D / 31 H |